    Калибровка нуля модуля CO2 модели MD ET7

    Июнь 2026 2 min read

    Обнуление модуля CO2 модели MD-ET7 Модуль CO2 модели MD-ET7 совместим только с аксессуарами CO2 бокового потока National Medical

    Zeroing the Model MD-ET7 CO2 module
    The Model MD-ET7 CO2 module is compatible only with National Medical’s side stream
    CO2 accessories. The Model MD-ET7 CO2 module can automatically inspect if the ZERO
    being correct or not and run an offset calibration program if the ZERO is error. When module runs the offset calibration program, there is no waveform being displayed (about 6 seconds), when the program running down, CO2 waveform will display again.
    Also the operator can run an offset calibration if he feels the zero is error, but we DO NOT suggest an operator to do so. An “Offset calibration” is a quick process that allows the Host system to adjust to the special characteristics of a particular Model MD-ET7 CO2 module; it is necessary only when requested. Such a request will occur if a change is detected in the Model MD-ET7 CO2 module.
     System does not allow offset calibration for 20 seconds after the last breath is detected.
     System does not allow offset calibration if temperature is not stable
     Offset calibration program maybe get a wrong ZERO if the ZERO hole is occluded.

    NOTE
    DO NOT jam the ZERO hole at any time. For best results, wait 5 minutes to allow the
    Model MD-ET7 CO2 module to warm up before performing the offset calibration procedure.

    Directions for Use of SINGLE PATIENT USE Nasal Side stream Kit
    This kit is intended for monitoring the CO2 of non-incubated patients. 
    Select the sidestream annular kit that is appropriate for the patient size and application.

    CAUTION:
    The Nasal Cannula kit are intended for single patient use. Do NOT reuse or sterilize the cannula kit as system performance will be compromised.
    1. Verify that the cannula kit is clean, dry and undamaged. Replace the cannula kit if necessary.
    2. Insert the gas drier line into the Gas In connector as shown in Figure 1 on previous
    page. Connect the gas drier line and the gas sampling line as shown in Figure 1. Connect

    Insert the gas drier line into the Gas In connector as shown in Figure 1 on previous page. Connect the gas drier line and the gas sampling line as shown in Figure 1
    the gas sampling line and the Nasal Cannula as shown in  :
     http://www.meditech.com.cn/Meditech-Library/How-to-connect-Meditech-Capnography-monitor-with-Anesthesia-machine—   
    3. Place the nasal cannula kits onto the patient as shown in Figure 2. 

    Place the nasal cannula kits onto the patient as shown in Figure 2.
    Share:

    Больше из раздела «Новости и события»

    Смотреть все статьи
    Библиотека Meditech
    Библиотека Meditech Авг 2026

    Таблица сравнения фетальных мониторов MD901F и MD901F-Android

      Параметр MD901F MD901F-Android Тип 12.1 дюйм 12.1 дюйм Стандартная комплектация FHR, TOCO, FM FHR, TOCO, FM, сенсорный экран, сеть…

    Читать далее 2 min
    библиотека продуктов
    библиотека продуктов Июн 2026

    Спирометр iBreathe с сенсорным экраном и встроенным принтером

    СПИРОМЕТР iBreathe С СЕНСОРНЫМ ЭКРАНОМ со встроенным принтером Портативное устройство для тестирования функции легких Этот спирометр использует технологию дифференциального давления…

    Читать далее 2 min
    библиотека продуктов
    библиотека продуктов Июн 2026

    Как производятся электроды для АВД?

    Как производятся электроды для АВД Производство электродов для АВД — это тщательный процесс, включающий несколько этапов для обеспечения их эффективности…

    Читать далее 2 min